Grid 2100

The JPI Mammography Grid is
A Carbon-Interspaced and Carbon-fiber covered grid which is optimal for a digital mammography unit.
Using the same cutting process as semi-conductor manufacturers, the JPI Grid-2100 can be controlled to micrometer precision, which eliminates the effects of non-uniform densities such as Moire phenomenon.
Specifications
| Materials | Carbon-Interspaced with Carbon-Fiber Cover |
| Dimentions | 18x24cm or 24x30cm |
| Densities | 50 LPC-78LPC |
| Ratios | 5:1 or 6:1 |
| FD | 60cm or 65cm |

Physical Carachteristics
- Test results
1. Increased Transmission - Improved by 9%
2. Lower dosage - Bucky Factor Reduced by 24%
3. Higher Sharpness of image - Selectivity Improved by 12%

- The test followed the clause #4, the measurement and determination of physical characteristics, of
the IEC 627 Standard. (100kvp, 200mA) - Test measurements are controlled under Gauge R&R, the key requirement of the Six Sigma.
- The test results can be slightly different due to test specimen and test conditions.